The Yokogawa EJA510E high-performance absolute pressure transmitter is a high-performance direct insertion threaded absolute pressure transmitter among the E-series products of Chongqing Chuan Instrument Yokogawa transmitters. It adopts the latest single crystal silicon resonant sensor technology, mainly suitable for pressure measurement of liquids, gases, or steam. The Yokogawa EJA510E high-performance absolute pressure transmitter converts the measured pressure into a 4-20mA DC current signal output, featuring fast response, remote setting, self-diagnosis, and other functions. The Yokogawa EJA E-series transmitters provide communication protocols such as BRAIN, HART/HART (1~5VDC) low power type, FF fieldbus, and PROFIBUS PA, with standard configuration passing SIL 2 safety certification. The Yokogawa EJA-E series transmitters are the latest DPharp series launched by Yokogawa Electric Corporation in Japan. It was released in 2012, combining the robustness and success of the Yokogawa EJA-E series with the performance of the pure EJX-A series to provide the type of products you expect from Yokogawa.
Product specifications of the Yokogawa EJA510E high-performance absolute pressure transmitter:
| Parameter | Specification |
|---|---|
| Model | EJA510E |
| Measured Variable | Absolute Pressure |
| Accuracy | ±0.055% of span (standard), ±0.04% (optional) |
| Stability | ±0.2% of URL over 10 years |
| Response Time | 90 milliseconds |
| Range Limits | Up to 21 MPa (3045 psi) |
| Overpressure Limit | Up to 8,700 psia |
| Output Protocols | HART 5/7, FOUNDATION Fieldbus, PROFIBUS PA, BRAIN, Low Power HART |
| Sensor Type | DPharp Digital Resonant Sensor |
| Safety Certifications | SIL 2/3 (TÜV and Exida certified) |
| Display | Optional digital LCD with bar graph |
| Local Configuration | Local Parameter Setting (LPS) available |
| Mounting Style | In-line mount |
| Power Supply | 10.5–42 V DC (dependent on protocol) |
| Ambient Temperature Range | -40 to +85°C |
Functional features of the Yokogawa EJA510E high-performance absolute pressure transmitter:
Absolutely—here are the key features of the Yokogawa EJA510E High-Performance Absolute Pressure Transmitter:
High Accuracy: Standard ±0.055% of span; optional enhanced accuracy of ±0.04%
Excellent Long-Term Stability: ±0.2% of URL over 10 years
Fast Response Time: As quick as 90 ms for rapid process changes
Wide Measurement Range: Up to 21 MPa (about 3045 psi)
Overpressure Durability: Withstands up to 8,700 psia
Safety Certifications: SIL 2/3 (TÜV and Exida certified) for use in safety instrumented systems
DPharp Digital Sensor: Superior reliability and signal clarity vs. analog sensors
Multiple Protocol Support: HART 5/7, FOUNDATION Fieldbus, PROFIBUS PA, BRAIN, and low-power HART
Local Parameter Setting (LPS): Enables configuration without external tools
Optional Display: Easy-to-read digital LCD with bar graph and indicator icons
Advanced Diagnostics: Proactive alerts for maintenance and troubleshooting
Selection specifications and common models of the Yokogawa EJA510E high-performance absolute pressure transmitter:
model | Specification code | explain |
EJA510E | Pressure transmitter | |
output signal | - D | 4-20mA DC BRAN protocol |
- J | 4-20mA DC HART 5/HART7 protocol | |
- F | Refer to GS01C31T02-01CN for FF fieldbus protocol | |
- G | Refer to GS01C31T02-01CN for the PROFIBUSPA bus protocol | |
- Q | 1-15V DC low-power HART7 protocol | |
Measurement range (Capsule) | A | 10~200kPa(1.45~29Psi) |
B | 0.1~2MPa(14.5~290Psi) | |
C | 0.5~10MPa(72.5~1450Psi) | |
D | 5~50MPa(720~7200Psi) | |
Material of the liquid receiving part | Process joint membrane, other | |
S | 316L SST Hastelloy C-276 316LSST | |
H | Hastelloy C-276 Hastelloy C-276 Hastelloy C-276 | |
Process connection | 4 | 1/2NPT internal thread |
7 | 1/2NPT external thread | |
8 | G1/2 DIN 16288 external thread | |
9 | M20X1.5 DIN 16288 external thread | |
---- | N | Usually N |
---- | -0 | Usually 0 |
Amplifier casing | 1 | Cast aluminum alloy |
3 | Corrosion resistant cast aluminum alloy | |
2 | ASTM CF-8M stainless steel | |
Electrical connection | 0 | G1/2 internal thread, one electrical interface without blind plug |
2 | 1/2NPT internal thread, two electrical interfaces without blind plugs | |
4 | M20 internal thread, two electrical interfaces without blind plugs | |
5 | G1/2 internal thread, two electrical interfaces with a blind plug | |
7 | 1/2NPT internal thread, two electrical interfaces with a blind plug | |
9 | M20 internal thread, two electrical interfaces with a blind plug | |
A | G1/2 internal thread, two electrical interfaces with one SUS316 blind plug | |
C | 1/2NPT internal thread, two electrical interfaces with one SUS316 blind plug | |
D | M20 internal thread, two electrical interfaces with one SUS316 blind plug | |
Built in display table | D | Digital display table |
E | Digital display table with range setting switch | |
N | nothing | |
2-inch pipeline installation bracket | L | SUS316 |
N | No installation bracket |
Friendly reminder:Due to the numerous selection parameters of the Yokogawa EJA510E high-performance absolute pressure transmitter, if you need a complete product selection and model quotation, please feel free to call for detailed inquiries.
Frequently Asked Questions – Yokogawa EJA510E
1. What is the EJA510E used for?
It’s designed to measure absolute pressure in industrial processes—ideal for vacuum systems, sealed vessels, and high-pressure applications where atmospheric pressure changes must be excluded.
2. What makes the EJA510E different from other pressure transmitters?
It uses Yokogawa’s DPharp digital resonant sensor, which offers better long-term stability, faster response, and higher accuracy than traditional analog sensors.
3. What communication protocols does it support?
It supports HART 5/7, FOUNDATION Fieldbus, PROFIBUS PA, BRAIN, and Low Power HART, making it highly versatile for integration into various control systems.
4. Is it suitable for safety-critical applications?
Yes, it’s SIL 2/3 certified by TÜV and Exida, making it suitable for use in Safety Instrumented Systems (SIS).
5. Can I configure it without a handheld communicator?
Absolutely. It features Local Parameter Setting (LPS), allowing basic configuration directly from the device using pushbuttons.
6. What are the available pressure ranges?
It offers multiple capsule options, covering ranges from 10 kPa to 50 MPa (1.45 to 7200 psi), depending on the selected model variant.
7. What materials are available for wetted parts?
You can choose from 316L stainless steel or Hastelloy C-276, depending on your process compatibility needs.
8. Is a digital display available?
Yes,
Common models:
| Model Code | Capsule | Material (Wetted Parts) | Electrical Connection | Mounting / Options |
|---|---|---|---|---|
| EJA510E-DAS4N-09DE | D (MPa capsule) | A (316L SS Diaphragm) | S (G1/2 female) | 4N (HART 5/7, LPS, LCD); 09DE |
| EJA510E-EAS4N-09DE | E (Higher range capsule) | A (316L SS Diaphragm) | S (G1/2 female) | Same as above |
| EJA510E-DBS4N-09DE | D | B (Hastelloy C-276) | S (G1/2 female) | Same as above |
| EJA510E-EBS4N-09DE | E | B (Hastelloy C-276) | S (G1/2 female) | Same as above |
| EJA510E-DCS4N-09DE | D | C (Tantalum) | S (G1/2 female) | Same as above |
| EJA510E-ECS4N-09DE | E | C (Tantalum) | S (G1/2 female) | Same as above |
| EJA510E-DDS4N-09DE | D | D (Gold-plated Diaphragm) | S (G1/2 female) | Same as above |
| EJA510E-EDS4N-09DE | E | D (Gold-plated Diaphragm) | S (G1/2 female) | Same as above |